ChemWell 2 Automated Analyzers
ChemWell 2全自動分析儀
產(chǎn)品說明
Specifications | |
Fluid handling | Pipetting range 5 μL – 1.95 mL Typical reaction volume range of 200 μL to 250 μL |
Probes | Liquid sensing probes monitor remaining volumes, crash protection |
ELISA reader | Accuracy of ±1% + 0.01 A or better up to 1.5 A and±2% or better above 1.5 A |
ELISA Optical filter range | 4 standard filters 405, 450, 492, 630 nm, optional additional filters available |
CLIA reader | Glow Luminescence |
Sensitivity/Detection limit | HRP 1x10-18 moles, ALP 1x10-21 moles |
Detector | Photomultiplier (PMT) |
Spectral Response Range | 300-650 nm |
Peak Wavelength | 400 nm |
Light source | Tungsten-Xenon lamp |
Standard sample rack | 90 sample capacity - serum vials or 13x100 mm tubes |
Standard reagent rack | Typically 29 - Assorted replaceable racks and custom designedracks are available for various bottle sizes |
Barcode scanner | Integrated barcode scanning with positive patient ID system |
Dimensions | Height - 57.7 cm Width - 107 cm Depth - 62.4 cm Weight- 60 kg |